SCPI Command Reference

:SENSe:SP[1-1]:FREQuency:SPAN

frequency offset)
When the downconverter is turned on and the RF input is set to 'E5052A
Direct'(Allowable setting range varies depending on whether the frequency offset is
on or off , the LO frequency within the frequency offset, and the conversion mode
used in the frequency offset)
When the downconverter is turned on and the RF input is set to 'Downconverter'
Equivalent key
SP Menu -> Start/Center -> Center
SP Menu -> Stop/Span -> Center
SP Menu -> Marker To -> Marker -> Center
:SENSe:SP[1-1]:FREQuency:SPAN
Syntax
:SENSe:SP[1-1]:FREQuency:SPAN <numeric>
:SENSe:SP[1-1]:FREQuency:SPAN?
Description
Sets/Reads the span value of frequency span
Parameter
When the E5052A is used stand-alone, or with the downconverter turned off, or with
the downconverter on and the RF input is set to 'E5052A Direct'(Allowable setting
range varies depending on whether the frequency offset is on or off , the LO
316
<Numeric>
Range
10M to 117G
Preset value
1G
Unit
Hz
Resolution
100m
<Description>
Range
10M to 113G
Preset value
3G
Unit
Hz
Resolution
100m
<Description>
Range
3G to 26.5G
Preset value
3G
Unit
Hz
Resolution
100m
7