Parameter
Range
Preset value
Unit
Resolution
Equivalent key
No equivalent key is available on the front panel.
:STATus:OPERation:CONDition
Syntax
:STATus:OPERation:CONDition
Description
Reads operation status conditional register value (Query Only)
Equivalent key
No equivalent key is available on the front panel.
:STATus:OPERation:ENABle
Syntax
:STATus:OPERation:ENABle <numeric>
:STATus:OPERation:ENABle?
Description
Sets/Reads operation status enable register
Parameter
Range
Preset value
Unit
Resolution
Equivalent key
No equivalent key is available on the front panel.
:STATus:OPERation[:EVENt]
Syntax
:STATus:OPERation[:EVENt]
Description
Reads operation status event register (Query Only)
Equivalent key
No equivalent key is available on the front panel.
7
<Numeric>
0 to 14
-
-
-
<Numeric>
0 to 65535
0
-
-
SCPI Command Reference
:STATus:OPERation:CONDition
349