SCPI Command Reference

:STATus:QUEStionable:POWer:ENABle

:STATus:QUEStionable:POWer:ENABle
Syntax
:STATus:QUEStionable:POWer:ENABle <numeric>
:STATus:QUEStionable:POWer:ENABle?
Description
Sets/Reads questionable-power status enable register
Parameter
Equivalent key
No equivalent key is available on the front panel.

:STATus:QUEStionable:POWer[:EVENt]

Syntax
:STATus:QUEStionable:POWer[:EVENt]
Description
Reads questionable-power status event register value (Query Only)
Equivalent key
No equivalent key is available on the front panel.

:STATus:QUEStionable:PTRansition

Syntax
:STATus:QUEStionable:PTRansition <numeric>
:STATus:QUEStionable:PTRansition?
Description
Sets/Reads questionable status positive transition filter value
Parameter
Equivalent key
No equivalent key is available on the front panel.
364
<Numeric>
Range
0 to 65535
Preset value
0
Unit
-
Resolution
-
<Numeric>
Range
0 to 65535
Preset value
32767
Unit
-
Resolution
-
7