Source:voltage:power:level:state; Source:voltage:power:limit:high - Agilent Technologies E5052A Programmer's Manual

Signal source analyzer
Hide thumbs Also See for E5052A:
Table of Contents

Advertisement

Resolution
Equivalent key
PN Menu -> DC Power Voltage -> DC Power Voltage
SP Menu -> DC Power Voltage -> DC Power Voltage
FP Menu -> DC Power Voltage -> DC Power Voltage
TR Menu -> DC Power Voltage -> DC Power Voltage
USER Menu -> DC Power Voltage -> DC Power Voltage

:SOURce:VOLTage:POWer:LEVel:STATe

Syntax
:SOURce:VOLTage:POWer:LEVel:STATe {ON|OFF|1|0}
:SOURce:VOLTage:POWer:LEVel:STATe?
Description
Truns on/off the DC Power output voltage
Parameter
ON or 1
OFF or 0(Preset
value)
Equivalent key
PN Menu -> DC Power Voltage -> DC Power Output
SP Menu -> DC Power Voltage -> DC Power Output
FP Menu -> DC Power Voltage -> DC Power Output
TR Menu -> DC Power Voltage -> DC Power Output
USER Menu -> DC Power Voltage -> DC Power Output

:SOURce:VOLTage:POWer:LIMit:HIGH

Syntax
:SOURce:VOLTage:POWer:LIMit:HIGH <numeric>
:SOURce:VOLTage:POWer:LIMit:HIGH?
Description
Sets/Reads maximum voltage limit for DC Power
Parameter
Range
Preset value
7
:SOURce:VOLTage:POWer:LEVel:STATe
<Numeric>
1m
Description
Set DC Power voltage output to 'ON'
Set DC Power volage output to 'OFF'
<Numeric>
0 to 16
16
SCPI Command Reference