:DISPlay:MESSage:CLEar

Syntax
:DISPlay:MESSage:CLEar
Description
Clears caution/message (No Query)
Equivalent key
No equivalent key is available on the front panel.

:DISPlay:PN[1-1]:ALLTrace:PERSistence:CLEar

Syntax
:DISPlay:PN[1-1]:ALLTrace:PERSistence:CLEar
Description
Clears all persistent traces (No Query)
Equivalent key
No equivalent key is available on the front panel.

:DISPlay:PN[1-1]:ANNotation:MARKer:POSition

Syntax
:DISPlay:PN[1-1]:ANNotation:MARKer:POSition {LEFT|RIGHt}
:DISPlay:PN[1-1]:ANNotation:MARKer:POSition?
Description
Sets/Reads the marker information position
Parameter
LEFT
RIGHt(Preset
value)
Equivalent key
PN Menu -> Display -> Marker Information

:DISPlay:PN[1-1]:ANNotation:MEASurement:STATe

Syntax
:DISPlay:PN[1-1]:ANNotation:MEASurement:STATe {ON|OFF|1|0}
:DISPlay:PN[1-1]:ANNotation:MEASurement:STATe?
Description
Turns on/off measurement conditions
Parameter
ON or 1(Preset
value)
OFF or 0
7
Description
Set the marker information position to 'Left'
Set the marker information position to 'Right'
Description
Set measurement conditions mode to 'ON'
Set measurement conditions mode to 'OFF'