Calculate:tr[1-1]:Trace[1-4]:Hold; Calculate:tr[1-1]:Trace[1-4]:Limit:fail; Calculate:tr[1-1]:Trace[1-4]:Limit:lower:ldata - Agilent Technologies E5052A Programmer's Manual

Signal source analyzer
Hide thumbs Also See for E5052A:
Table of Contents

Advertisement

SCPI Command Reference

:CALCulate:TR[1-1]:TRACe[1-4]:HOLD

Equivalent key
TR Menu -> Marker Function -> Analysis Type
:CALCulate:TR[1-1]:TRACe[1-4]:HOLD
Syntax
:CALCulate:TR[1-1]:TRACe[1-4]:HOLD {OFF|MAXimum|MINimum}
:CALCulate:TR[1-1]:TRACe[1-4]:HOLD?
Description
Sets/Reads trace data hold type
Parameter
Equivalent key
TR Menu -> Trace View -> Data Hold

:CALCulate:TR[1-1]:TRACe[1-4]:LIMit:FAIL

Syntax
:CALCulate:TR[1-1]:TRACe[1-4]:LIMit:FAIL
Description
Reads out the limit test result (Query Only)
Parameter
When the limit test is set to OFF, OFF or 0 is always read out.
Equivalent key
No equivalent key is available on the front panel.

:CALCulate:TR[1-1]:TRACe[1-4]:LIMit:LOWer:LDATa

Syntax
:CALCulate:TR[1-1]:TRACe[1-4]:LIMit:LOWer:LDATa <array>
:CALCulate:TR[1-1]:TRACe[1-4]:LIMit:LOWer:LDATa?
186
Description
STATistics
Set analysis type to 'Statistics'
Description
OFF(Preset value)
Set data hold to 'Off'
MAXimum
Set data hold to 'Maximum'
MINimum
Set data hold to 'Minimum'
Description
ON or 1
The limit test result is fail
OFF or 0(Preset
The limit test result is pass
value)