Sense:sp[1-1]:Carrier:fband; Sense:sp[1-1]:Carrier:set:center - Agilent Technologies E5052A Programmer's Manual

Signal source analyzer
Hide thumbs Also See for E5052A:
Table of Contents

Advertisement

SCPI Command Reference

:SENSe:SP[1-1]:CARRier:FBANd

Equivalent key
SP Menu -> Average/BW -> RBW
:SENSe:SP[1-1]:CARRier:FBANd
Syntax
:SENSe:SP[1-1]:CARRier:FBANd {LOW|HIGH}
:SENSe:SP[1-1]:CARRier:FBANd?
Description
Sets/Reads the carrier frequency band
Parameter
When the E5052A is used stand-alone, or with the downconverter turned off, or with
the downconverter on and the RF input is set to 'E5052A Direct'
When the downconverter is turned on and the RF input is set to 'Downconverter'
Equivalent key
SP Menu -> Start/Center -> Carrier To -> Frequency Band
SP Menu -> Stop/Span -> Carrier To -> Frequency Band

:SENSe:SP[1-1]:CARRier:SET:CENTer

Syntax
:SENSe:SP[1-1]:CARRier:SET:CENTer <numeric>
Description
Changes the center frequency to N times the carrier frequency (No Query)
Parameter
314
<Numeric>
Preset value
25k
Unit
Hz
Resolution
-
Description
LOW
Set the carrier frequency band to '10M - 1.5GHz'
HIGH(Preset
Set the carrier frequency band to '300M - 7GHz'
value)
<Description>
BAND3(Preset
Set the carrier frequency band to '3G - 7GHz'
value)
BAND4
Set the carrier frequency band to '9G - 26.5GHz'
<Numeric>
Range
1 to 20
7