SCPI Command Reference

:FORMat:DATA

Parameter
Equivalent key
No equivalent key is available on the front panel.
:FORMat:DATA
Syntax
:FORMat:DATA {ASCii|REAL32|REAL64}
:FORMat:DATA?
Description
Sets/Reads data transfer mode
Parameter
Equivalent key
No equivalent key is available on the front panel.

:HCOPy:ABORt

Syntax
:HCOPy:ABORt
Description
Aborts printing (No Query)
Equivalent key
PN Menu -> System -> Abort Printing
SP Menu -> System -> Abort Printing
FP Menu -> System -> Abort Printing
TR Menu -> System -> Abort Printing
USER Menu -> System -> Abort Printing

:HCOPy:IMAGe

Syntax
:HCOPy:IMAGe {NORMal|INVert}
:HCOPy:IMAGe?
272
Description
NORMal(Preset
Set byte order so that a byte containing MSB (Most Significant
value)
Bit) is transferred first
SWAPped
Set byte order so that a byte containing LSB (Least Significant
Bit) is transferred first
Description
ASCii(Preset
Set data transfer mode to 'ASCii'
value)
REAL32
Set data transfer mode to 'IEEE 32 floating point binay'
REAL64
Set data transfer mode to 'IEEE 64 floating point binay'