Display:user[1-1]:Trace[1-8]:Limit:line; Display:user[1-1]:Trace[1-8]:Mode; Display:user[1-1]:Trace[1-8]:Persistence:state - Agilent Technologies E5052A Programmer's Manual

Signal source analyzer
Hide thumbs Also See for E5052A:
Table of Contents

Advertisement

SCPI Command Reference

:DISPlay:USER[1-1]:TRACe[1-8]:LIMit:LINE

Equivalent key
USER Menu -> Trace View -> Trace Label
:DISPlay:USER[1-1]:TRACe[1-8]:LIMit:LINE
Syntax
:DISPlay:USER[1-1]:TRACe[1-8]:LIMit:LINE {ON|OFF|1|0}
:DISPlay:USER[1-1]:TRACe[1-8]:LIMit:LINE?
Description
Turns on/off the limit line
Parameter
Equivalent key
USER Menu -> Display -> Limit Test -> Limit Line

:DISPlay:USER[1-1]:TRACe[1-8]:MODE

Syntax
:DISPlay:USER[1-1]:TRACe[1-8]:MODE {OFF|DATA|MEMory|BOTH}
:DISPlay:USER[1-1]:TRACe[1-8]:MODE?
Description
Sets/Reads data and/or memory trace display
Parameter
Equivalent key
USER Menu -> Trace View -> Display Trace

:DISPlay:USER[1-1]:TRACe[1-8]:PERSistence:STATe

Syntax
:DISPlay:USER[1-1]:TRACe[1-8]:PERSistence:STATe {ON|OFF|1|0}
:DISPlay:USER[1-1]:TRACe[1-8]:PERSistence:STATe?
Description
Clears persistent data
266
Description
ON or 1(Preset
Turn on the limit line mode
value)
OFF or 0
Turn off the limit line mode
Description
OFF
Set data and/or memory trace to 'Off'
DATA(Preset
Set data and/or memory trace to 'Data'
value)
MEMory
Set data and/or memory trace to 'Memory'
BOTH
Set data and/or memory race to 'Both' (data and memory)