:TRIGger:FP[1-1]:MODE

*1
Syntax
:TRIGger:FP[1-1]:MODE {ANALyzer|TESTer}
:TRIGger:FP[1-1]:MODE?
Description
Selects measurement mode in the Frequency. power and DC currnet measurement mode
Parameter
ANALyzer(Preset
value)
TESTer
Equivalent key
FP Menu -> Trigger -> Mode

:TRIGger:FP[1-1]:SOURce

Syntax
:TRIGger:FP[1-1]:SOURce {INTernal|EXTernal|MANual|BUS}
:TRIGger:FP[1-1]:SOURce?
Description
Selects trigger source
Parameter
INTernal(Preset
value)
*1."Option not installed" error message is generated when setting the trigger mode to the analyzer mode with the
option 011 instrument.
7
Description
Set External trigger polarity to 'NEGative'
Set External trigger polarity to 'POSitive'
Description
Set trigger mode to 'ANALyzer'
Set trigger mode to 'TESTer'
Description
Set trigger source to 'INTernal'
SCPI Command Reference
:TRIGger:FP[1-1]:MODE
371