Sense:tr[1-1]:Narrow:time:span; Sense:tr[1-1]:Power:input:level:maximum - Agilent Technologies E5052A Programmer's Manual

Signal source analyzer
Hide thumbs Also See for E5052A:
Table of Contents

Advertisement

SCPI Command Reference

:SENSe:TR[1-1]:NARRow:TIME:SPAN

Description
Selects reference position for time span of transient measurement
Parameter
Equivalent key
TR Menu -> Time Offset -> Narrow Ref Position
TR Menu -> Span -> Narrow Ref Position
:SENSe:TR[1-1]:NARRow:TIME:SPAN
Syntax
:SENSe:TR[1-1]:NARRow:TIME:SPAN <numeric>
:SENSe:TR[1-1]:NARRow:TIME:SPAN?
Description
Sets/Reads time span
Parameter
Equivalent key
TR Menu -> Time Offset -> Narrow Span
TR Menu -> Span -> Narrow Span

:SENSe:TR[1-1]:POWer:INPut:LEVel:MAXimum

Syntax
:SENSe:TR[1-1]:POWer:INPut:LEVel:MAXimum <numeric>
:SENSe:TR[1-1]:POWer:INPut:LEVel:MAXimum?
Description
Sets/Reads the maximum input level for transient measurement
Parameter
324
Description
LEFT
Set reference position to 'LEFT'
CENTer(Preset
Set reference position to 'CENTer'
value)
RIGHt
Set reference position to 'RIGHt'
<Numeric>
Range
0 to 100m
Preset value
100m
Unit
s
Resolution
10n
<Numeric>
Range
-45 to 30
7