Calculate:user[1-1]:Trace[1-8]:Limit:lower:segment:data; Calculate:user[1-1]:Trace[1-8]:Limit:report[:Data]; Calculate:user[1-1]:Trace[1-8]:Limit[:State] - Agilent Technologies E5052A Programmer's Manual

Signal source analyzer
Hide thumbs Also See for E5052A:
Table of Contents

Advertisement

Unit
Resolution
Equivalent key
No equivalent key is available on the front panel.
:CALCulate:USER[1-1]:TRACe[1-8]:LIMit:LOWer:SEGM
ent:DATA
Syntax
:CALCulate:USER[1-1]:TRACe[1-8]:LIMit:LOWer:SEGMent:DATA <array>

:CALCulate:USER[1-1]:TRACe[1-8]:LIMit:LOWer:SEGMent:DATA?

Description
Sets/Reads segment data of the lower limit line
Parameter
Range
Preset value
Unit
Resolution
Equivalent key
No equivalent key is available on the front panel.
:CALCulate:USER[1-1]:TRACe[1-8]:LIMit:REPort[:DATA
]
Syntax

:CALCulate:USER[1-1]:TRACe[1-8]:LIMit:REPort[:DATA]

Description
Reads the limit test results of all measurement points in selected traces (Query Only)
Equivalent key
No equivalent key is available on the front panel.

:CALCulate:USER[1-1]:TRACe[1-8]:LIMit[:STATe]

Syntax
:CALCulate:USER[1-1]:TRACe[1-8]:LIMit[:STATe] {ON|OFF|1|0}
:CALCulate:USER[1-1]:TRACe[1-8]:LIMit[:STATe]?
Description
Turns on/off the limit test function