… Warning Level Setting

Memo
The warning level is set to "80%" when the instrument is shipped from the factory.
Operating Procedure
1
Use [▲] or [▼] to move the cursor to "Warning
level", and then press the [Confirmation] key.
The <Warning level> screen is displayed.
2
Use [▲] or [▼] to change the value.
Settings
{ 000 to 100%
3
Press the [Confirmation] key after completing the
changes.
The setting is confirmed and the screen returns to the
<Default data setup> screen.
Notes
If [ESC] is pressed during configuration, the settings will
not be changed and the screen will return to the
<Default data setup> screen.
75
Start the procedure from the <Setting> - <Default data setup> screen.