X85; Specular Component - Konica Minolta CM-23d Instruction Manual

Table of Contents

Advertisement

… Specular Component

Select the method for processing the sample specular reflection light.
Memo
The specular component is set to "SCI + SCE" when the instrument is shipped from the factory.
Operating Procedure
1
Use [▲] or [▼] to move the cursor to "Specular
Component", and then press the [Confirmation] key.
2
Use [▲] or [▼] to move the cursor to the item to set,
and then press the [Confirmation] key.
Settings
{ SCI
: Measurement using a light source that
contains specular reflection light.
{ SCE
: Measurement using a light source that does
not contain specular reflection light.
{ SCI+SCE : Displays the results of both SCI and SCE
measurements.
3
Press the [Confirmation] key.
The selection is confirmed and the screen returns to the
<Measurement setup> screen.
Notes
If [ESC] is pressed without pressing the [Confirmation]
key, the settings will not be changed and the screen will
return to the <Measurement setup> screen.
Start the procedure from the <Measurement cond.> - <Measurement setup> screen.