Konica Minolta CM-23d Instruction Manual page 137

Table of Contents

Advertisement

Symptom
Data cannot be
Is the USB cable connected
transferred to a computer
correctly?
from the instrument.
No commands from the
Has Bluetooth communication
computer are being
been established?
accepted.
Printing is not possible.
Has Bluetooth communication
been established?
Samples and settings are
The instrument's backup
not held in memory and
battery may be low
disappear immediately.
immediately after purchase or
following a period of
prolonged non-use. Turn the
instrument ON to charge the
backup battery. Under this
condition, the backup battery
will be fully charged in
20 hours.
Check Point
Properly connect the instrument's USB connecting
terminal to the computer's USB port with the USB
cable supplied with the instrument.
Install and connect the Bluetooth module correctly.
Install and connect the Bluetooth module correctly.
The backup battery in the instrument should last
10 years under normal use. If the instrument fails to
retain data in memory even after the battery has been
fully charged, however, the battery has likely reached
the end of its life and requires replacement.
To replace the backup battery, please contact a
KONICA MINOLTA-authorized service facility.
Action
5
134