… Parametric Coefficient Setting

Specify the parametric coefficient used for pass/fail judgment of a sample for each target.
Operating Procedure
1
Use [▲] or [▼] to move the cursor to "Parametric
coef.", and then press the [Confirmation] key.
The <Parametric coef.> screen is displayed.
2
Use [▲] or [▼] to move the cursor to the item to set,
and then press the [Confirmation] key.
The <Parametric coef.> edit screen is displayed.
3
Use [▲] or [▼] to move the cursor to the item to set,
and press the [Confirmation] key to change the
setting.
• Pressing the [Confirmation] key causes the cursor to
appear on the value. Press [▲] or [▼] to change the
value. Move the cursor between the digits of the value
by pressing [◀] or [▶].
Settings
{ 0.01 to 9.99
• Press the [Confirmation] key to confirm every time the
setting of an item is changed.
4
After all settings have been configured, use [▲] or
[▼] to move the cursor to "OK", and then press the
[Confirmation] key.
The setting is confirmed and the screen returns to the
<Pass/Fail> screen.
Memo
If SCI+SCE is set for "Specular Component" under
<Measurement setup>, use [◀] or [▶] to switch between
SCI and SCE.
Notes
If [ESC] is pressed during configuration, the settings will
not be changed and the screen will return to the <Pass/
Fail> screen.
5
Press [ESC] to return to the <Pass/Fail> screen.
Start the procedure from the <Target menu> - <Pass/Fail> screen.