… Zero calibration skip

When shipped from the factory, the "Calibration (Excluding Zero Cal.)" button is enabled on the <Calibration>
screen. If configuring to ensure that zero calibration is performed, disable this function.
Memo
Zero calculation skippable is set to "Disable (Perform zero calibration)" when the instrument is shipped from the
factory.
Notes
When using the MAV target mask (with glass), it is necessary to set "Zero cal. skippable" to "Disable" because
the amount of stray light tends to fluctuate.
Operating Procedure
1
Use [▲] or [▼] to move the cursor to "Zero cal.
skippable", and then press the [Confirmation] key.
2
Use [▲] or [▼] to move the cursor to select
"Disable" or "Enable".
3
Press the [Confirmation] key.
The selection is confirmed and the screen returns to the
<Calibration setup> screen.
Notes
If [ESC] is pressed without pressing the [Confirmation]
key, the settings will not be changed and the screen will
return to the <Calibration setup> screen.
Start the procedure from the <Setting> - <Calibration setup> screen.