MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) MODEL GFC7000EU ANALYZER (Addendum to GFC7000E Manual PN 04584) Teledyne Analytical Instruments 16830 Chestnut St. City of Industry, Ca. 91748 TEL: 626-961-9221 FAX: 626-961-2538 TEL: 626-934-1500 E-MAIL: ai-customerservice@teledyne.com WEB SITE: http://www.teledyne-ai.com...
(Addendum to GFC7000E Manual PN 04584) OVERVIEW The model GFC7000EU is a close derivative of the Model GFC7000E CO Analyzer. The optical bench is longer (14 m), making the instrument more sensitive at low CO levels. The instrument has a user selectable full scale range of 0-100 PPB to 0-100 PPM of CO The instrument does not come equipped with a sample flow rate sensor.
“From Purge” is the exhaust port for the for the purge gas. The gas needs to exhausted with a gas line not more than 10 meters long. In addition, the GFC7000EU uses a different sample pressure sensor than the GFC7000E. The GFc7000EU pressure sensor is rated for oxygen service. The part number is listed in the appendix –...
Page 4
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) There are 4 relay alarm outputs (AL1-AL4) on the rear panel. AL1 is for system okay, AL2 is for concentration limit 1 exceeded and AL3 is for concentration limit 2 exceeded.
Page 5
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) SAMPLE WHEEL TEMP WARNING CO = XX.XX TEST deactivates Warning Messages until New warning(s) TEST SETUP are activated MSG activates Warning SAMPLE RANGE=50.0 PPM CO = XX.XX Messages.
Page 6
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) Warning Fault Condition Possible Causes Message FRONT PANEL The CPU is unable to WARNING only appears on Serial I/O COM Port(s) WARN Communicate with Front Panel Display will be frozen, blank or will not the Front Panel respond.
Relay controlling the Wheel Heater Entire Relay Board I C Bus APPENDIX B – GFC7000EU Spare Parts List NOTE Use of replacement parts other than those supplied by API may result in non-compliance with European standard EN 61010-1.