To set up the external trigger test

Performance Test only
This test verifies that the Agilent 35670A meets its trigger specification for external
trigger level accuracy. In this test, a frequency synthesizer outputs a 1kHz signal to
the external trigger input and a 12.8 kHz signal to channel 1. The analyzer makes
an accurate triggered measurement on channel 1 to verify the trigger level and
slope.
1
2 ch
Using an HP 3326A
1
4 ch
Using an HP 3326A
Verifying Specifications
To set up the external trigger test