Chapter 6: Replacing Assemblies

To remove and replace the AutoProbe assembly

To remove and replace the AutoProbe assembly

Use this procedure to remove and replace the AutoProbe assembly. When necessary, refer to
other removal procedures.
Figure 6-4
AutoProbe Assembly
Disconnect the power cable and remove the cover.
1
Remove the AutoProbe connector assembly, the subpanel, and the probe connector
2
assembly that fits around the front-panel BNC connectors, by doing the following:
a Disconnect the mylar ribbon cable W8 from the Probe Power and Control Board, A15.
Figure 6-5
Disconnecting W13
6–6
See "To disconnect and connect Mylar flex cables" in this chapter.