Output Pod Characteristics - Agilent Technologies 16522A User Manual

200-m vectors/s pattern generator
Table of Contents

Advertisement

Agilent Technologies 16522A Reference

Output Pod Characteristics

Output Pod Characteristics

The following equivalent circuit information is provided to help you select the
appropriate clock and data pods for your application.
Agilent 10461A TTL Data Pod
Output type
Maximum clock
Skew
Recommended lead set
Agilent 10462A 3-State TTL/CMOS Data Pod
Output type
3-state enable
Maximum clock
Skew
Recommended lead set
2–18
10H125 with 100 ohm in series
200 MHz
Typical <2 ns; worst case 4 ns (note 1)
Agilent 10474A
74ACT11244 with 100 ohm in series
10H125 on non 3-state channel 7 (note 2),
negative true, 100K ohm to GND,
enabled on no connect
100 MHz
Typical <4 ns; worst case 12 ns (note 1)
Agilent 10474A

Advertisement