Selecting The Module - Agilent Technologies 16522A User Manual

200-m vectors/s pattern generator
Table of Contents

Advertisement

Programming the Agilent 16522A

Selecting the Module

Program Comments
Line 10 selects the pattern generator in slot A
Line 20 removes all labels previously assigned
Line 30 assigns label 'A', positive polarity and assigns the seven least
significant bits of pod A5
Line 40 assigns label 'B' and assigns all eight bits of pod A4
Line 50 removes all program lines
Line 60 inserts a new line (after line 0) in the INIT SEQUENCE portion of the
program.
Line 70 inserts a new line (after line 4) in the MAIN SEQUENCE portion of
the program. Recall that the default MAIN SEQUENCE already has two lines
of program
Line 80 Sets the RMODE to repetitive. If the program is to be run only once,
select the :RMODE SINGLE command.
Line 90 Starts the program.

Selecting the Module

Before you can program the pattern generator, you must first "select" it,
otherwise, there is no way to direct your commands to the pattern generator.
To select the module, use the system command :SELect , followed by the
numeric reference for the slot location of the pattern generator(1...10 refers
to slot A...J respectively). For example, if the pattern generator master card
is in slot E, then the command:
:SELect
5
would select this module. For more information on the SELect command,
refer to the Agilent 16500 Programmer's Guide.