Installation

To install modules

To install modules

Slide the cards which are above the slots into which you want to put
1
modules about halfway out of the mainframe.
With the probe cables facing away from the instrument, slide the
2
modules approximately halfway into the mainframe.
Slide the assembled modules into the mainframe, but not completely
3
in.
Each card in the instrument will be firmly seated and tightened in step 5.
Position all cards and filler panels so that the endplates overlap.
4
Seat the cards and tighten the thumbscrews.
5
Starting with the bottom card, firmly seat the cards into the backplane
connector of the mainframe. Keep applying pressure to the center of the
card endplate while tightening the thumbscrews finger-tight. Repeat this for
all cards and filler panels from the bottom up.
Correct air circulation keeps the instrument from overheating. For correct
C A U T I O N
air circulation, filler panels must be installed in all unused card slots. Keep
any extra filler panels for future use.