To Determine Peak To Average Characteristics - Agilent Technologies E8247C PSG CW User Manual

Psg signal generators
Table of Contents

Advertisement

To Determine Peak to Average Characteristics

This procedure describes how to set the phases of the tones in a multitone waveform and determine the peak
to average characteristics by plotting the complementary cumulative distribution function (CCDF).
Mode > Multitone > Initialize Table
1. Press
Freq Spacing
2. Press
>
Initialize Phase Fixed Random
3. Press
Done
4. Press
.
Apply Multitone
5. Press
More (1 of 2) > Waveform Statistics > Plot CCDF
6. Press
You should now see a display that is similar to the one shown in
peak to average characteristics of the waveform with all phases set to zero.
Figure 8-7
Peak
Power
Mode Setup > Initialize Table
7. Press
Initialize Phase Fixed Random
8. Press
Random Seed Fixed Random
9. Press
Chapter 8
Number of Tones
>
20
kHz
>
.
to Fixed.
.
CCDF Plot with Fixed Phase Set
.
to Random.
to Random.
Creating, Viewing, and Optimizing Multitone Waveforms
64
Enter
>
>
.
.
Figure
Multitone Waveform Generator
8-7. The CCDF plot displays the
177

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8257c psgE8267c psgE8267cE8257cE8247c

Table of Contents