Servicing the Agilent 16522A

Testing Interval
Periodic performance verification (full calibration) is not required for
the Agilent 16522A module. Because there are no specifications for
the Agilent 16522A, testing against specifications does not apply.
To Use the Flowcharts
Flowcharts are the primary tool used to isolate defective assemblies.
The flowcharts refer to other tests to help isolate the trouble. The
circled letters on the charts indicate connections with the other
flowcharts. Start your troubleshooting at the top of the first flowchart.
With Agilent 16500B Mainframe
Before starting the troubleshooting, ensure that operating system
v3.05 or later is installed on the mainframe. To check the operating
system version number, enter the System Test menu.
If v3.05 or later is not loaded, obtain a copy of the latest operating
system software and install it on the logic analyzer hard drive. After
installation, recycle power on the mainframe.
With Agilent 16500C Mainframe
The Agilent 16522A works with all operating system versions on the
Agilent 16500C.