How The Transmission Measurement Works - HP 8712ES User Manual

Rf network analyzers
Hide thumbs Also See for HP 8712ES:
Table of Contents

Advertisement

Figure 3-36
Figure 3-37
Impedance Calculation for Reflection Measurements

How the Transmission Measurement Works

In a transmission measurement, the data can be converted to its
equivalent mathematical series impedance using the model and
equations shown in
Impedance Calculation for Transmission Measurements
In the formula shown above, T is the complex transmission response.
The complex impedance, Z
analyzer displays the magnitude of Z
two-port Z-parameter conversion, as only the measured parameter is
used in the equations.
Chapter 6
provides information about all of the calibrations available for
transmission measurements. A two-port calibration is the most accurate
form of error correction for transmission measurements since it accounts
for all of the major sources of systematic error. Another possible
calibration choice for a transmission measurement is the enhanced
response calibration. However, since the enhanced response calibration
ES User's Guide
Figure
3-37.
, is computed based on T and Z
Trans
Trans
Making Measurements
Measuring Impedance Magnitude
. This is not the same as a
. The