HP 8712ES User Manual page 450

Rf network analyzers
Hide thumbs Also See for HP 8712ES:
Table of Contents

Advertisement

Table 9-9
General Information (continued)
Description
System Bandwidths
Wide (6500 Hz)
Medium Wide (4000 Hz)
Medium (3700 Hz)
Medium Narrow (1200 Hz)
Narrow (250 Hz)
Fine (15 Hz)
Rear Panel
Auxiliary Input:
Connector
Impedance
Range
Accuracy
Damage Level
External Trigger In/Out:
Damage Level
Limit Test Output:
Damage Level
User TTL Input/Output:
Damage Level
Specifications
HP 8712ET/ES and HP 8714ET/ES General Information
±10 V
±3% (reading) + 20 mV
>+15 V; <−15 V
<−0.2 V; >+5.2 V
<−0.2 V; >+5.2 V
<−0.2 V; >+5.2 V
9-40
Specification
Supplemental Information
6500 Hz, nom
4000 Hz, nom
3700 Hz, nom
1200 Hz, nom
250 Hz, nom
15 Hz, nom
Female BNC
10 kΩ, nom.
Female BNC; Open-collector
with 681 Ω nom. pullup resistor
to +5 V, nom. Normally high,
pulsed low after each data point
is measured.
Female BNC; Open-collector
with 681 Ω nom. pullup resistor
to +5 V, nom. Normally high,
pulled low when limit test fails.
Female BNC; Open-collector
with 681 Ω nom. pullup resistor
to +5 V nom. Programmable as:
high-sweep output; trigger
input; general I/O for IBASIC.
ES User's Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 8714es