Making Measurements
Measuring S11 Reflection Port 1 using a One-Port Calibration
To perform a one-port calibration you will need one of the following
calibration kits depending on the nominal impedance of your analyzer:
Impedance
75 Ω
75 Ω
75 Ω
50 Ω
50 Ω
50 Ω
50 Ω
50 Ω
50 Ω
50 Ω
By convention, cal kit labels in the analyzer indicate the sex of the port
with which they are used. For example, the default cal kit for the
analyzer is type-N female because the front panel RF ports are female
(the calibration standards, in turn, have male connectors).
If you are going to be using calibration standards other than the default
(female type-N), you must select a calibration kit to match the port(s) at
your calibration reference plane. Do this by pressing
Cal Kit
and then selecting the appropriate type.
3-50
For Use
Connector
with Male
Type
Test Ports
type-F
type-N
type-N
type-N
type-N
APC-7
N/A
3.5 mm
7-16
7-16
7-16
For Use
with Female
Test Ports
N/A
More Cal
CAL
ES User's Guide