HP 8712ES User Manual page 389

Rf network analyzers
Hide thumbs Also See for HP 8712ES:
Table of Contents

Advertisement

Softkey
Auto-Step
Space
Span
Split Disp
FULL split
Spur Avoid
Softkey in the user TTL config menu. Displays a pop-up message with
a description of softkey sequencing and softkeys to enable or disable
the auto-step function.
See "Measurement Setup and Control with Fast Recall" in the
Automating Measurements User's Guide Supplement for more
information.
Access Keys:
SYSTEM OPTIONS
User TTL Config
Softkey in the character entry menu. Adds a blank space to a title or
filename.
Softkey in the
FREQ
Selects the frequency span of the source. When selected, it changes the
frequency annotation from start/stop to center/span.
See
"Entering Measurement Parameters" on page 2-4
information.
Softkey in the more display menu. Toggles the display mode between
split (measurement channel 1 on top, measurement channel 2 on the
bottom) and full (both measurement channels on full screen).
See
"Entering Measurement Parameters" on page 2-4
information.
Access Keys:
DISPLAY
Softkey in the spur avoid options menu. If this option is selected, the
analyzer switches between the default and an alternate configuration
during a sweep to avoid spurs.
NOTE
See
"Reducing Trace Noise" on page 5-12
CAUTION
Access Keys:
MENU
ES User's Guide
System Config
menu. Used in conjunction with
More Display
Sweeptime usually increases when this option is
selected.
The measurement calibration must be performed
with the same spur avoid option used in the
measurement or your results may be invalid.
Spur Avoid Options
Hardkey/Softkey Reference
Center
for more
for more
for more information.
S
.
8-83