HP 8712ES User Manual page 238

Rf network analyzers
Hide thumbs Also See for HP 8712ES:
Table of Contents

Advertisement

Activating Spur
Avoidance
NOTE
CAUTION
CAUTION
Optimizing Measurements
Reducing Trace Noise
When you activate spur avoidance, the analyzer sweeps to a point before
a spur, stops the sweep, shifts the spur, sweeps through the spur
location, then shifts the spur back and continues the sweep. The
analyzer determines which spurs need to be avoided with an algorithm
based on frequencies, number of points, sweep time, and system
bandwidth.
To activate spur avoidance:
1. Press
MENU
2. Make a user-defined measurement calibration. For calibration
procedures, refer to
Using spur avoid increases sweep time. Since there are more spurs at the
lower frequencies, the time penalty can be reduced by setting the start
frequency of the measurement as high as possible.
The measurement calibration must be performed with the same spur
avoid mode used in the measurement or your results may be invalid.
You will invalidate the measurement calibration if you turn spur avoid
off.
5-14
Spur Avoid Options
Chapter
6.
Spur Avoid
.
ES User's Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 8714es

Table of Contents