Table 9-8
Test Port Input (continued)
System Dynamic Range
HP 8712ES:
50 Ω:
Narrowband
Broadband (Internal)
75 Ω:
Narrowband
Broadband (Internal)
HP 8714ES:
50 Ω:
Narrowband
Broadband (Internal)
75 Ω:
Narrowband
Broadband (Internal)
a. The System Dynamic Range is calculated as the difference between the receiver noise floor and
the minimum of either the source maximum output (maximum power setting minus output
power level accuracy) or the receiver maximum input. System Dynamic Range applies to trans-
mission measurements only, since reflection measurements are limited by directivity. The Sys-
tem Dynamic Range for HP 8714 ET/ES 75 Ω analyzers is not a specification for frequencies >2
GHz; it is a characteristic. For the HP 8712ES/8714ES, external broadband detectors will pro-
vide more dynamic range than the internal broadband detectors.
HP 8712ET/ES and HP 8714ET/ES Test Port Input
Description
a
(continued)
ES User's Guide
Spec.
(in dB)
System Bandwidths:
Fine
Fine
(15 Hz)
(15 Hz)
104
115
46
53
99
111
39
47
101
112
43
50
96
108
36
44
Specifications
Typical
(in dB)
Med Wide
Wide
(4000 Hz)
(6500 Hz)
90
57
46
33
87
54
40
27
87
54
43
30
84
51
37
24