Numeric Entries

Markers number one through eight in the
any one of these softkeys makes the marker the active marker and (if
previously off) turns it on. ">" after the marker number means the
marker is the active marker. ":" means the marker is not the active
marker; it may be on or off.
See
"Using Markers" on page 4-3
and for more information on using markers.
Softkey in the cal kit menu of 50 ohm instruments. Selects the type of
cal kit as 3.5 mm.
See
Chapter 6, "Calibrating for Increased Measurement Accuracy,"
more information.
Access Keys:
CAL
NOTE
Softkey in the system impedance menu. Selects 50 ohms as the system
impedance.
Refer to
"When to Change the System Impedance" on page 3-12
more information.
Access Keys:
CAL
ES User's Guide
for an explanation of "active marker,"
More Cal Cal Kit
By convention, cal kits indicate the sex of the port
with which they are used. For example, the
default cal kit for the analyzer is type-N female
because the front panel RF ports are female (the
calibration standards, in turn, have male
connectors).
More Cal System Z0
Hardkey/Softkey Reference
Numeric Entries
menu. Pressing
MARKER
for
for
8-3