Table 9-6
System Performance, Uncorrected, 75Ω
Description
c
Directivity
d
Source Match (Ratio)
e
Load Match
f
Crosstalk
Directivity
Source Match (Ratio)
Load Match
Reflection Tracking
Transmission Tracking
f
Crosstalk
a. Option 1E1 adds a 60 dB step attenuator to the HP 8712ET/8714ET.
b. These numbers apply for a measurement made using the "Fine" bandwidth at an environmental temperature of
25° ± 5° C.
c. The uncorrected directivity of a network analyzer is calculated in linear terms by dividing the reflection measurement
of a perfect load by the average of the reflection measurements of an ideal short and an ideal open.
d. The uncorrected source match is the source match of the network analyzer when making a ratioed, uncalibrated mea-
surement.
e. The uncorrected load match is the match of the network analyzer port used on the load side of a measurement.
f.
Measured by setting output power to the maximum specified setting, connecting shorts to both ports, and measuring
transmission. Typical and specified crosstalk values are 5 dB worse than those shown in the table below 1 MHz (for all
models) and above 2.2 GHz (for 8714ET/ES models).
Specifications
HP 8712ET/ES and HP 8714ET/ES (75 Ω)
HP 8712ES/8714ES
300 kHz
1.3 GHz
to 1.3 GHz
to 3 GHz
Specification
27
19
18
15
18
15
88
88
Typical
40
30
23
20
24
22
±
±
1.5
1.5
±
±
1.5
1.5
95
95
9-22
HP 8712ET/8714ET
with Attenuator
a
Option 1E1
300 kHz
1.3 GHz
to 1.3 GHz
to 3 GHz
b
(in dB)
27
19
21
15
18
15
97
97
b
(in dB)
40
30
24
20
24
22
±
±
1.0
1.5
±
±
1.5
1.5
105
105
HP 8712ET/8714ET
(without Attenuator)
300 kHz
1.3 GHz
to 1.3 GHz
to 3 GHz
27
18
23
19
18
15
97
97
40
30
28
23
24
22
±
±
1.0
1.0
±
±
1.0
1.0
105
105