Cal Kit Model
Number
HP 85039B
HP 85036B
HP 85036E
HP 85032B
HP 85032E
HP 85031B
HP 85033D
HP 85038A
HP 85038F
HP 85038M
NOTE
Impedance
75 Ω
75 Ω
75 Ω
50 Ω
50 Ω
50 Ω
50 Ω
50 Ω
50 Ω
50 Ω
By convention, cal kit labels in the analyzer indicate the sex of the port
with which they are used. For example, the default cal kit for the
analyzer is type-N female because the front panel RF ports are female
(the calibration standards, in turn, have male connectors).
Calibrating for Insertable Devices
When measuring a through-standard during calibration, the test ports
normally mate directly together. For example, two cables with the
appropriate connectors can be connected directly without a
through-adapter, resulting in a zero-length through-standard. An
insertable device is a device that can be inserted in place of a zero-length
through-standard. It has the same type of connector on each port, but
with a different sex, or the same type of sexless connector on each port
(APC-7, for example).
ES User's Guide
Measuring S-Parameters using a Two-Port Calibration
For Use
Connector
with Male
Type
Test Ports
type-F
type-N
type-N
type-N
type-N
APC-7
N/A
3.5 mm
7-16
7-16
7-16
Making Measurements
For Use
with Female
Test Ports
N/A
3-21