HP 8712ES User Manual page 73

Rf network analyzers
Hide thumbs Also See for HP 8712ES:
Table of Contents

Advertisement

Cal Kit Model
Number
HP 85039B
HP 85036B
HP 85036E
HP 85032B
HP 85032E
HP 85031B
HP 85033D
HP 85038A
HP 85038F
HP 85038M
NOTE
Impedance
75 Ω
75 Ω
75 Ω
50 Ω
50 Ω
50 Ω
50 Ω
50 Ω
50 Ω
50 Ω
By convention, cal kit labels in the analyzer indicate the sex of the port
with which they are used. For example, the default cal kit for the
analyzer is type-N female because the front panel RF ports are female
(the calibration standards, in turn, have male connectors).
Calibrating for Insertable Devices
When measuring a through-standard during calibration, the test ports
normally mate directly together. For example, two cables with the
appropriate connectors can be connected directly without a
through-adapter, resulting in a zero-length through-standard. An
insertable device is a device that can be inserted in place of a zero-length
through-standard. It has the same type of connector on each port, but
with a different sex, or the same type of sexless connector on each port
(APC-7, for example).
ES User's Guide
Measuring S-Parameters using a Two-Port Calibration
For Use
Connector
with Male
Type
Test Ports
type-F
type-N
type-N
type-N
type-N
APC-7
N/A
3.5 mm
7-16
7-16
7-16
Making Measurements
For Use
with Female
Test Ports
N/A
3-21