NOTE
Figure 3-16
Measuring S21 Forward Transmission using an Enhanced Response Calibration
4. The analyzer will measure each standard and then calculate the new
calibration coefficients. The message "Calibration complete."
will appear for a few seconds when the analyzer is done calculating
the new error correction array.
Changing sweep frequencies (and other source parameters) may affect
your calibration. See
Calibrating for Noninsertable Devices
A non-insertable device is one that cannot be inserted in place of a
zero-length through-standard. It has the same connectors on each port
(type and sex) or has a different type of connector on each port (SMA on
one port, and Type-N on the other, for example). To review two methods
of calibration for non-insertable devices, refer to the previous section
"Measuring S-Parameters using a Two-Port Calibration" on page
both methods, you will need to select
calibration menu instead of

Connect the DUT

Equipment Setup for an S
ES User's Guide
Chapter 6
for more information.
Enhanced Response
User 2-Port
Measurement
21
Making Measurements
from the
.
3-19. In
3-45