HP 8712ES User Manual page 393

Rf network analyzers
Hide thumbs Also See for HP 8712ES:
Table of Contents

Advertisement

Switching
Test Set
SWR
Sync Green
on OFF
System
Bandwidth
System BW
Wide
Medium Wide (default)
Medium
Medium Narrow
Narrow
Fine
Softkey in the system config menu. Configures the analyzer for use
with a multiport test set.
Refer to the multiport test set User's and Service Guide for more
information.
Access Keys:
SYSTEM OPTIONS
Softkey in the
FORMAT
(standing wave ratio).
Softkey in the CRT adjust menu. Toggle to ON for use with a
sync-on-green external monitor. The default setting is toggled to OFF.
Access Keys:
SYSTEM OPTIONS
Softkey in the
AVG
the IF filters in the receiver section of the network analyzer.
See
"To Reduce the Receiver Noise Floor" on page 5-10
following table for more information on system bandwidth.
Dynamic Range
Low
Medium
Medium
Medium
High
High
ES User's Guide
System Config
menu. Displays the data formatted as SWR
System Config
menu. Displays the menu to set the bandwidth of
Sweep Speed
Fastest
Faster
Fast
Slow
Slower
Slowest
Hardkey/Softkey Reference
CRT Adjust
and the
Approximate BW
6500 Hz
4000 Hz
3700 Hz
1200 Hz
250 Hz
15 Hz
8-87
S