HP 8712ES User Manual page 95

Rf network analyzers
Hide thumbs Also See for HP 8712ES:
Table of Contents

Advertisement

Cal Kit Model
Number
HP 85039B
HP 85036B
HP 85036E
HP 85032B
HP 85032E
HP 85031B
HP 85033D
HP 85038A
HP 85038F
HP 85038M
NOTE
NOTE
Measuring S21 Forward Transmission using an Enhanced Response Calibration
To perform an enhanced transmission response calibration, you will need
one of the following calibration kits depending on the nominal impedance
of your analyzer:
Impedance
75 Ω
75 Ω
75 Ω
50 Ω
50 Ω
50 Ω
50 Ω
50 Ω
50 Ω
50 Ω
By convention, cal kit labels in the analyzer indicate the sex of the port
with which they are used. For example, the default cal kit for the
analyzer is type-N female because the front panel RF ports are female
(the calibration standards, in turn, have male connectors).
If you are going to be using calibration standards other than the default
(female type-N), you must select a calibration kit to match the port(s) at
your calibration reference plane. Do this by pressing
Cal Kit
and then selecting the appropriate type.
ES User's Guide
For Use
Connector
with Male
Type
Test Ports
type-F
type-N
type-N
type-N
type-N
APC-7
N/A
3.5 mm
7-16
7-16
7-16
Making Measurements
For Use
with Female
Test Ports
N/A
More Cal
CAL
3-43

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 8714es

Table of Contents