23. When the instrument prompts you, connect three standards (open,
short, and load) to the port 1 test cable as shown in the graphic on the
following page.
24. Press
Measure Standard
25. When the instrument prompts you, connect three standards (open,
short, and load) to the port 2 test cable—as shown in the graphic
above.
26. Press
Measure Standard
27. The analyzer will measure each standard and then calculate new
calibration coefficients. The message "Calibration complete."
will appear for a few seconds when the analyzer is done calculating
the new error correction array.
ES User's Guide
Measuring S-Parameters using a Two-Port Calibration
after connecting each standard.
Port 2: Open, Short, Load Connections
after connecting each standard.
Making Measurements
3-33