NOTE
CAUTION
All of the above color settings are retained when the analyzer is turned
off or is preset. To restore the default colors, press
More Display Color Options

Synchronizing and Positioning the Display

The analyzer provides a CRT adjustment feature which can be used to
get an external monitor to synchronize properly, and to optimize the
display's position. The following list explains how to use the CRT adjust
features.
The CRT adjust settings also affect the analyzer's built-in display.
1. Press
SYSTEM OPTIONS
this key is pressed, a test pattern appears on the monitor(s). If you
don't want the test pattern on the display while adjusting it, press
Remove Pattern
2. If your external monitor is a sync-on-green-signal monitor, press the
Sync Green on OFF
3. Use the
Vertical Position
vary the display's position on the monitor. After pressing one of these
softkeys, use the front panel knob, the
numeric keypad to vary the position of the display on the monitor.
4. Press
More
. For synchronizing your external monitor, use the
Vertical Back Porch
Vertical Frnt Porch
the vertical and horizontal front and back porch times until the
display is optimal. Use
5. All of these settings are retained when the analyzer is turned off or
preset.
ES User's Guide
,
Factory Default
System Config
.
key to enable the sync-on-green capability.
and
Horizontal Position
,
Horizontal Back Porch
, and
Horizontal Frnt Porch
Restore Defaults