Set 4: WKSTN83A (C = 1 through 8)
The linear in density curves in this ULUT Set use contrast settings of 1 through 8 (Figure 2-10).
The best curve (Contrast setting) to obtain a linear in density response depends on the current Density
(Dmax) setting, as indicated in the following table.
Density Setting
When the Contrast setting is selected from the table for this ULUT Set, the density of the print should
decrease in approximately equal steps of Optical Density as the Digital Input is increased by a given step
size.
2004 December Rev.S 1908110
Figure 2-10.
15,16
13,14
11,12
9,10
7,8
5,6
3,4
1,2
INSTALLATION INSTRUCTIONS
Contrast Setting for
"Linear in Density" Response
1
2
3
4
5
6
7
8
2-43