Kodak DryView 8300 Service Manual page 320

Laser imager
Hide thumbs Also See for DryView 8300:
Table of Contents

Advertisement

SERVICE MANUAL
All Parameters Can Now Be Sent to the Service Port: By sending a [CTRL] P all the imager
parameters are sent to the service PC for capture and can be saved on a diskette for future reference.
Modalities Can Continue to Try to Send Images in Target Mode When Memory Full: The user then
does not have to press the Enter button to clear the memory full message on the local panel.
The SCSI Target Contrast Test Now Uses Same Memory Full Error Message: This is now consistent
with all other 8300 imager software.
Version 4.6 (Effective 6/30/99, beginning with Serial Number 1900)
False PROCESSOR DRUM Related Error Codes (EC 17, 19, 88): Eliminated.
Configure Client Defaults: Sub-Menu item was restored (was lost as a bug in Version 4.5).
FAI Response: Will Be Sent to Host when CLI fails, but an ALM 61 will no longer be sent.
Version 4.5
There is a bug in this version, so it was never released. All of these features have been added to Version
4.6, with the bug corrected.
Calibration Sheet: Now Rebuilt in Memory each time a print is made so any corruption is cleared.
Ethernet Default Router Address: Now an Ethernet installation menu item.
SCSI Target: Now Reports Sense Data for errors that are related to the modality for that target.
False "Unable to Read DIR Entry" Bernoulli Disk Error: Eliminated.
Regulating Temperature: The Laser will no longer stick in the Regulating Temperature.
SCSI Target Mode: No longer implements optional "FRU" which is a sense data field.
Host Control RES Command: Now cancels any contrast test prints requested by CNT but not
processed.
SCSI Target Mode: Displays Contrast & Density in the "System Configuration" of the "Test Menu".
Input B Image too Large to Print: Displays if SCSI DISK image is too large by pixels, etc.
SCSI Target Mode: Now makes a print of the entire Image, including the last line pixel and last image
line.
Image DRAM Diagnostic: Now Correctly Counts the number of bad blocks of image memory.
Setting a Suspended Modality Back to Ready: Now prevented by completing an acquisition.
Duplicated Code: Duplicated DISK formats for ATL and Beroulli have been eliminated.
Power On Self Test: Now done on the local buffer used to pass data to the ASIC.
Magnification Factor of Less than 1.0: Will never be applied to an Aegis image that is too large.
AIM Command from Ethernet Software: Will no longer be rejected by the SYSTEM software.
SCSI Initiator Image Pixel and Line Replication: Has been corrected and will make a print of a whole
image.
"LAMP Failed to Reach Temperature" Error: Now gives more brightness information to the SERVICE
PORT.
"Trailing Blank Images": Now recognized as a host control film with no visible images ignored.
Processing the Host Reset (RES) Command: Will now flush the output queue of any old responses.
1908110 2003 September Rev. Q