Kodak DryView 8300 Service Manual page 40

Laser imager
Hide thumbs Also See for DryView 8300:
Table of Contents

Advertisement

SERVICE MANUAL
Parameter
11. CAL PROCESSOR
CENTER TEMP
1. MEASURED CENTER
ZONE READING
12. CAL PROCESSOR END
TEMP
1. MEASURED END ZONE
READING
13. ENERGY SAVER OPTION
14. IMAGE VIEWING
17. PM EXPOSURES
18. FILM COUNT ENABLE
(effective with Version 4.9)
2-22
Options
None
None
ON or OFF
EMULSION or AH (anti-halation)
Set to 0 at installation.
ON or OFF
Action
Do not adjust at installation.
Current temperature is displayed
and can range from 112.8 --
140.6°F. This parameter is used
in temperature calibration
procedure (paragraph 3-1).
Same as above.
Leave at OFF. If set to ON,
HEATERS and MOTORS will
de-energize after 3 hours if the
IMAGER is not used. This option
is not programmable.
Leave at EMULSION (except
when necessary to change).
Reset to 0 when PM has been
completed. "SCHEDULE PM"
message will be removed from
LOCAL PANEL.
Set to ON to allow the number of
remaining films in the FILM
TRAY to be displayed on the
LOCAL PANEL.
1908110
2004 December Rev. S