Kodak DryView 8300 Service Manual page 59

Laser imager
Hide thumbs Also See for DryView 8300:
Table of Contents

Advertisement

Set 1: UPMSV3K
This ULUT Set consists of a more complete set of "saggy" curves, providing a wider choice of
brightnesses than curves 1--6 of the U1_109B ULUT Set (see Figure 2-6). This set is widely used in
applications including ultrasound, nuclear medicine, and others where an "S-shaped" response is not
desired. Because this ULUT Set is suitable for most ultrasound and nuclear medicine applications, it is an
appropriate choice for a majority of IMAGER users.
For these curves, overall image brightness increases as the Contrast setting is increased from 1 to 12.
The brighter curves in this set are often preferred by radiologists working in ultrasound or nuclear
medicine.
Set 2: 693C0
This ULUT Set consists entirely of "S-shaped" curves with a mid-tone brightness that gives an
approximately "linear in perceived brightness" response, especially for the central Contrast settings of this
set (Figure 2-7).
This ULUT Set is suitable for CT, MR and C-Arm applications. It is also useful for some nuclear medicine
applications. Increasing the Contrast setting from 1 to 12 increases the mid-tone contrast and decreases
the contrast in the near-black and near-white portions of the image.
2004 December Rev.S 1908110
Figure 2-6.
Figure 2-7.
INSTALLATION INSTRUCTIONS
2-41