Check that there are no ignition sources within at least 3-5 feet of the DRUM. This includes, but is not
limited to, cigarettes, SWITCHES, ELECTRICAL OUTLETS, LIGHT BULBS, and flames.
After cleaning, place used CLEANING PADS in BAG (78-9998-2868-0). TIE--WRAP the BAG and
dispose of properly.
UPPER PRESSURE ROLLERS
1. Check the ROLLER ASSEMBLY and behind the ROLLERS. If these areas have "FAZ" on them,
remove the ROLLERS to clean them:
a. Remove the GROUNDING CLIP SCREW and ground WIRE (if necessary) from the lower SHAFT
on the left side.
b. On the right side of the COVER, remove the SCREWS (1 for each SHAFT) from the ends of the 2
SHAFTS connecting to the ROLLER ASSEMBLY.
c. Pull the SHAFTS out from the left side.
d. Remove the ROLLER ASSEMBLY.
e. Clean the metal surfaces exposed by removal of the ROLLERS with WIPES. Use
alcohol-dampened CLEANING PADS to remove dirt buildup.
Figure 5-10. Removing the UPPER PRESSURE ROLLER ASSEMBLY
!
Caution
To prevent "deplating" the ROLLERS, do not rub rub the ROLLERS too hard with the abrasive
CLEANING PADS.
2. Clean the ROLLERS with WIPES. Use CLEANING PADS to remove thick stains.
3. If deplating or degradation (such as flat spots) on the ROLLERS is visible, install a new ROLLER
ASSEMBLY.
4. If you removed the UPPER ROLLER ASSEMBLY from the PROCESSOR COVER, replace it after
cleaning. Installation is the reverse of disassembly (see step 1).
Note
After installing, check that the GROUNDING CLIP makes contact with the SCREW (See
Figure 5-11).
2004 December Rev. S 1908110
d
a
c
Tools/Preventive Maintenance/Cleaning
b
5-15