Kodak DryView 8300 Service Manual page 269

Laser imager
Hide thumbs Also See for DryView 8300:
Table of Contents

Advertisement

TROUBLESHOOTING
7-2-72. Maximum Density Less than 3.0
LOCAL PANEL Message
Host Alarm
Service Log
MAXIMUM DENSITY LESS THAN 3.0
No
Yes
SUMMARY
Film calibration has failed.
SEQUENCE OF EVENTS
The IMAGER will not pass the film calibration pass if D max is less than 3.0.
PHONE FIX - - OPERATOR CORRECTABLE
Try loading new film.
ON SITE - -TECHNICIAN CORRECTABLE
Check that the DENSITOMETER is clean. Check the expiration date on the customer's film. Check the
storage procedures the customer uses.
TECH TIPS
THEORY
2004 December Rev. S
1908110
7-101