SERVICE MANUAL
Set 3: 713C0
This ULUT Set is a family of "S-shaped" curves similar to the 693C0 curves, but with a somewhat
elevated brightness (see Figure 2-8).
This ULUT Set has wide applicability. It is often used for ultrasound, nuclear medicine, and DSA and
might also be used in other MODALITIES when a brightness greater than the 693C0 Set is desired.
Set 4: WKSTN83A
Most of the curves in this ULUT Set are designed to give a "linear in density" response, while the other
curves in this set are designed to give an approximate "linear in brightness" response. Because of its
linear in density curves, this ULUT Set is uniquely suitable for use with certain MODALITIES or
WORKSTATIONS (e.g., Toshiba nuclear medicine or some Siemens workstations) that are designed to
work with an IMAGER which has a linear in density characteristic. Overall, this ULUT Set offers the
choices seen in Figure 2-9.
2-42
Figure 2-8.
Figure 2-9.
1908110
2004 December Rev. S