SERVICE MANUAL
Parameter
11. CAL PROCESSOR CENTER
TEMP
1. MEASURED CENTER
ZONE READING
ZONE READING
12. CAL PROCESSOR END
TEMP
1. MEASURED END ZONE
READING
13. ENERGY SAVER OPTION
14. IMAGE VIEWING
16. MODALITY GROUP
17. PM EXPOSURES
18. FILM COUNT ENABLE
(effective with Version 4.9)
2-14
Options
None.
Do not adjust at installation. Current
temperature is displayed and can range
temperature is displayed and can range
from 112.8 -- 140.6° F. This parameter is
used in the Setting Up PROCESSOR
Temperature procedure.
None.
Same as above.
ON or OFF
Leave at OFF. If set to ON, HEATERS and
MOTORS de-energize after 3 hours if the
IMAGER is not used. This option is not
programmable.
EMULSION or AH
Leave at EMULSION (except as
(anti-halation)
necessary).
0 -- 4
Defaults to Group 0. (See paragraph 2-13.)
Set to 0 at
Reset to 0 when PM has been completed.
installation. Causes
"SCHEDULE PM" message will be removed
"SCHEDULE PM"
from the LOCAL PANEL.
to be displayed on
LOCAL PANEL
when 10,000 films
have been printed.
ON or OFF
Set to ON to allow the number of remaining
films in the FILM TRAY to be displayed on
the LOCAL PANEL.
Action