Kodak DryView 8300 Service Manual page 24

Laser imager
Hide thumbs Also See for DryView 8300:
Table of Contents

Advertisement

SERVICE MANUAL
Parameter
1. ASPECT RATIO
Note
If "3M protocol" is selected, options 1--4 of the Serial Control Interface are available.
3. SERIAL CONTROL
INTERFACE:
1. BAUD RATE
2. NUMBER OF DATA BITS
3. NUMBER OF STOP BITS 1 or 2
4. PARITY
5. DELIMITER
6. IGNORE LATE ACK/NAK
7. PROTOCOL
6. MARK INSTALL DATE
7. MEMORY MANAGEMENT
2-6
Table 2-2. Setting the Digital Installation Parameters
Options
0.50 to 2.00
1200, 2400, 4800,
9600, or 19200
7 or 8
ODD, EVEN, or
NONE
Comma or space.
ON or OFF.
3M (Host) or
KEYPAD
Date of install:
day/month/year
A = 4.0 MB
B = 4.0 MB
Action
Set to 1.0. After acquiring a first image,
changes to this value may be necessary.
Adjust this parameter to what is displayed on
the MONITOR of the MODALITY. If a circle or
square test pattern cannot be displayed on the
MONITOR, ask the operator to make a square
or rectangle with the characters from the
KEYBOARD of the MODALITY.
(Note: The "clocks per line" parameter must be
set correctly before you set up Aspect Ratio.)
Measure the sides of the square or rectangle
on the MONITOR and compare these values to
what is on the film. The ratio should be the
same. A value of .9 is normal for many
Ultrasound MODALITIES.
For settings:
Select between baud rates.
Select data bit settings for value.
Select stop bit setting for value.
Select parity setting between selections.
Select between options, for host requirements
to separate DZO commands.
Set to ON if the host device is a Kodak PACS
Link 9410 ACQUISITION SYSTEM.
Select the "Control Protocol" setting between
options.
If date is incorrect, change Date/Time
parameter in the "General Setup" menu.
Partition memory across 2 INPUT MODULES
as necessary. If using only 1 MODULE, all
memory should be set to that input. See
paragraph 2--6 for setting up memory.
1908110
2004 December Rev. S