Select 2m to configure your distance sensor to a range of 0.4 – 2m, or 10m, to
configure your distance sensor to a range of 0.4 – 10m.
The new configuration will be the default range until the next time you change it.
4.
DO
sensor calibration
2
If you need to calibrate a DO
enter the DO
calibration screen:
2
_ _ _ _ _
U s e t h e f o l l o w i n g
m e n u t o c a l i b r a t e
t h e D O 2 e l e c t r o d e .
S a l i n i t y : 0 p p t
Use the arrow buttons to select between 0, 5, 10, 15, 20, 25, 30 and 35 ppt and
press the Enter button to confirm and move to Calibrate altitude.
Use the arrow buttons to select between 0, 500, 1000, 1500, 2000, 2500, 3000
and 3500 ft and press the Enter button to confirm.
MultiLogPRO will display the calculated upper limit of the calibrated range and
will exit the DO
calibration screen.
2
The new calibration parameters will be saved until the next time you change
them.
5.
Screen contrast
Use the arrow buttons to adjust the LCD screen contrast.
Any contrast adjustment will be saved until the next time you change it.
6.
Connect to a computer or to a TI calculator
Press Forward
connect to a TI calculator.

6.2.10 Internal Clock and Calendar

The internal clock is set the first time you use the Setup command from the
WeatherLab software to program the MultiLogPRO, and is automatically
sensor press the Forward arrow button
2
D O 2 C A L _ _ _ _ _
to connect to a computer or the Backward
to
MultiLogPRO
to
77