3.
Click Add New Sensor.
4.
Type in the sensor name and a sensor unit.
5.
Type in two calibration values (two real values and the
corresponding output voltages of the sensor).
6.
Click OK.
WeatherLab adds the new sensor to the sensor list and sends the definitions to
MultiLogPRO.
5.5.5 Calibrate Sensor
1.
Connect the MultiLogPRO to the PC.
2.
Select the Calibrate Sensors option from the Logger menu.
3.
Choose a sensor from the Sensor drop list.
4.
Enter a distinct real value in each of the Real Value edit boxes and
the corresponding measured values in each of the Measured
Value edit boxes (The measured values are the values displayed
by WeatherLab when measuring the two real values.
48
WeatherLab